Detection and subtyping of Herpes simplex virus in clinical samples by LightCycler PCR, enzyme immunoassay and cell culture

HealthData.Gov

Updated Oct 29, 2026

About this dataset

Background Prompt laboratory diagnosis of Herpes simplex virus (HSV) infection facilitates patient management and possible initiation of antiviral therapy. In our laboratory, which receives various specimen types for detection of HSV, we use enzyme immunoassay (EIA) for rapid detection and culture of this virus. The culture of HSV has traditionally been accepted as the diagnostic 'gold standard'. In this study, we compared the use of real time PCR (LightCycler) for amplification, detection and subtyping of specific DNA with our in-house developed rapid and culture tests for HSV.

      Results
      The LightCycler PCR (LC-PCR) detected and subtyped HSV in 99% (66/67) of HSV positive specimens, compared to 81% (54/67) by rapid antigen EIA or 57% (36/63) by culture. A specimen was considered positive when two or more tests yielded HSV identifications or was culture positive. Discordant results were confirmed with an in-house developed PCR-ELISA or DNA sequence analysis. The typing results obtained with the LC-PCR and by culture amplified test were completely concordant.
    
    
      Conclusions
      This study showed that the LC-PCR provided a highly sensitive test for simultaneous detection and subtyping of HSV in a single reaction tube. In addition to increased sensitivity, the LightCycler PCR provided reduced turn-around-times (2 hours) when compared to enzyme immunoassay (4 hours) or culture (4 days).

Data and resources (1)

Prototype by Datopian · not an official government website · data from data.virginia.gov