Stable hepatitis C virus RNA detection by RT-PCR during four days storage

HealthData.Gov

Updated Oct 29, 2026

About this dataset

Background Suboptimal specimen processing and storage conditions of samples which contain hepatitis C virus (HCV) RNA may result in a decline of HCV RNA concentration or false-negative results in the detection of HCV RNA in serum. We evaluated the stability of HCV RNA in serum and clotted blood samples stored at room temperature or at 4°C for 4 days with the aim of optimizing the standard procedures of processing and storage of samples.

      Methods
      Blood from five HCV RNA positive patients was collected in tubes with and without separator gel, centrifuged 1 or 6 hours after collection. Samples were then left 6, 24, 48, 72 or 96 h at room temperature (21.5 – 25.4°C) or at 4°C before determining their HCV RNA level using the COBAS AMPLICOR HCV MONITOR Test, vs 2.0 (Roche Diagnostic Systems).
    
    
      Results
      The logarithm of the HCV RNA level measurements remained within a 0.3 value of the means for 4 days at both temperatures (room temperature or 4°C).
    
    
      Conclusions
      We conclude that blood samples may be collected and aliquoted within 6 h of collection and can be stored at 4°C for 72 hours as proposed by the manufacturer without significant differences in measured HCV RNA level. Our results indicate that lapses in this scheme may still yield reliable results.